The HIGH PRESURE DIFFUSER was designed for use with Lighthouse Airborne Particle Counters in applications where the sample is under pressure. Standard applications include sampling of CDA, Nitrogen and other inert gasses.
The Mini Multiplexer communicates via MODBUS/ASCII or MODBUS/RTU and was designed with two RJ-45 connectors to allow daisy-chaining with other RS-485 instruments. The Mini Multiplexer is compatible with SOLAIRs and other Lighthouse Instruments on a daisy-chain when all are running MODBUS/ ASCII.
Power, Analog Inputs and Digital Outputs are connected via screw terminals for easy field installation and service of the unit. Connection to a variety of sensors and devices is possible without special connectors or tools. The REMOTE Mounting Kit option provides additional flexibility for mounting theMini Multiplexer.
Process Qualification
Semiconductor Process Monitoring
Flat Panel Process Monitoring
Process Trend Analysis
Supports 0.1μm at 1 CFM
May be Used With Any Inert Gas
Stainless Steel Construction
Easy Set Up
Simple Installation
Fast Purge and Cleanup
Stainless Steel Construction
Output at Near Atmospheric Pressure
Small Footprint
Easy System Integration
International Sales and Service Support
Low Cost of Ownership
Specifications
Specifications
Description:
Isokinetic High Pressure Diffuser, Vented Version
Isokinetic High Pressure Diffuser, Ported Version
Materials:
Stainless Steel with Nitrile (Buna-N)
and Fluorocarbon (Viton) O-Rings
Stainless Steel with Nitrile (Buna-N)
and Fluorocarbon (Viton) O-Rings
Inlet Pressure Range:
30 to 150 psi [2 to 10.3 bar]
30 to 150 psi [2 to 10.3 bar]
Outlet Flow Rate:
0.1 CFM or 1 CFM
0.1 CFM or 1 CFM
Flow Control:
Metal Orifice
Metal Orifice
Isokinetic Sampling Point:
Adjustable over the Inlet Pressure Range
Adjustable over the Inlet Pressure Range
Sample Fluid:
Pressurized Air, Nitrogen and Inert Gases
Pressurized Air, Nitrogen and Inert Gases
Inlet and Outlet Ports:
1/4” Metal Gasket High Purity VCR Fitting
3/8” Compression, Union Fitting
1/4” Metal Gasket High Purity VCR Fitting
3/8” Compression, Union Fitting
Diameter:
1.31” [3.3 cm] at large cap
1.31” [3.3 cm] x 3.08” [7.8 cm]
Length:
8.14” [20.7 cm] at the 30 psi setting and
9.87” [25.1 cm] at the 150 psi setting
9.1” [23.1 cm] at the 30 psi setting and
10.83” [27.5 cm] at the 150 psi setting
Weight:
1.11 lbs [0.5 kg]
1.65 lbs [0.75 kg]
Cleanability:
Easily disassembled and cleaned;
replaceable o-rings
Easily disassembled and cleaned;
replaceable o-rings