Remote 5104V

With a sensitivity of 0.5 micron and a monitored flow rate of 1.0 CFM (28.3 LPM), the REMOTE 5104V provides real-time continuous data collection at a cost-effective price per point.

The REMOTE 5104V integrates seamlessly with large facility monitoring and building management systems. It transfers up to 4 channels of simultaneous particle count data using RS-485 Modbus.

  • VHP Compatible
  • Class 1 Division 2 Rating
  • Built-in Flow Monitoring
  • 0.5 – 5.0 μm (optional channels up to 25 μm)
  • 1.0 CFM (28.3 LPM) Flow Rate
  • Up to 4 Channels of Simultaneous Data
  • Meets ISO 21501-4 Standards
  • Real-time Continuous Monitoring
  • Small Size
  • 316L Stainless Steel, Wash Down Resistant/Splash-Tight Enclosure
  • Status Indicators
  • RS-485 Modbus
  • Interfaces to Facility Monitoring Systems
  • Uses External Vacuum Source
  • Designed for Reliability
  • 2 Year Warranty
  • International Suppor

  • Versatile Mounting Options
  • Easy to Install when Space is an Issue
  • Cost Effective Price per Point
  • Proven Data Integrity
  • Simple Integration into FMS Systems
  • Low Cost of Ownership

  • Glass Coating
  • Aseptic Processing
  • Isolators
  • Cleanrooms

Industries

  • Aerospace Industry
  • Pharmaceutical Industry
  • Biotechnology
  • Medical Applications

Specifications

Specifications

Size Range: 0.5 – 25.0 µm
Channel Sizes: Standard: 0.5, 5.0 µm
Optional 4 Channels: 0.5, 1.0, 5.0, 25.0 µm
Counting Efficiency: 50% @ 0.5μm; 100% for particles > 0.75μm (per ISO 21501-4)
Flow Rate: 1.0 CFM (28.3 LPM)
Laser Source: Laser Diode
Zero Count Level: < 1 count / 5 minutes (per ISO 21501-4)
Concentration Limits: 500,000 Particles/ft3 @ 5% Coincidence Loss
Data Storage Rotating Buffer, 2000 Records
Calibration: NIST Traceable
Vacuum Requirements: External Vacuum >18″ (45.72) of Hg
Air Flow: Internally Monitored
Communication Modes: RS-485/MODBUS
LED Indicators: Power, Flow, Service, Sampling
Connector: Industrial RJ45
Supporting Software: Lighthouse Monitoring System, LMS Express, LMS Express RT, LMS LMS Net
Enclosure: 316L Stainless Steel, Wash Down Resistant/Splash-Tight
Power: 24 VDC
Dimensions: 4.2″ (l) x 2.0″ (w) x 3.7″ (h) [10.7 x 5.08 x 9.398 cm]
Weight: 15.6 oz (0.44 kg)

Conditions

Conditions

Operating: 50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing