NanoCount 50+

Liquid Particle Counters

Get A Quote
NanoCount 50+

The NanoCount 50+ provides industry standard RS-485 MODBUS and 4-20mA communication protocols so integration into facility monitoring systems or use as a stand-alone device is quick and easy.

Using its integrated Ethernet TCP/IP interface allows the NanoCount 50+ to connect to standard corporate Ethernet LANs as easily as it connects to Facility Monitoring Systems.

The compact size of the NanoCount 50+ allows it to be installed where other liquid particle counters won’t fit, including point-of-use process control applications.

Incorporating the latest in laser optical particle counting technology and designed for reliability, theNanoCount 50+ is intended for continuous trouble-free operation.

  • Channel Sizes – 50nm, 100nm, 150nm, 200nm
  • Four Channels of Simultaneous Count Data
  • 100 ml per Minute Flow Rate
  • Built-In 3-1/2″ Touch Screen Display
  • Easily Integrates with Existing Building Automation and Factory Management Systems
  • 316L Stainless Steel Construction
  • Industry Standard RS-485 / MODBUS, 4-20mA and Ethernet Interfaces (Sealed Connectors)
  • NEMA Style Enclosure
  • Integrated Manual Flow Control with Digital Readout on Touch Screen Display

  • 2 Year Warranty
  • Cost Effective Price per Point
  • Easy System Integration
  • International Sales and Service Support
  • Low Cost of Ownership

  • Deionized Water Monitoring
  • Process Qualification and Monitoring
  • Aqueous Parts Cleaning

Specifications

Specifications

Size Range: 50 – 200nm
Channel Sizes: Standard: 50nm, 100nm, 150nm, 200nm
Flow Rate: 100 ml/min
Zero Count Level: < 20 counts/Liter
Laser Source: Laser Diode
Calibration: NIST Traceable
Communication Modes: RS-485 and Ethernet use Industrial RJ45s and 4-20mA uses sealed DB9
LED Indicators: Status
User Interface 3-1/2 inch TFT Touch Screen
Concentration Limit: 200,000 counts/ml @ 5% Coincidence Error
Enclosure: Stainless Steel
Sample Inlet/Outlet Connection: 1/4″ Flaretek™
Sample Temperature: 32 – 140°F (0 – 60°C)
Sample Pressure: 75 PSI
Wetted Surface Materials: Quartz, PTFE, PFA, Viton, Kel-F, Kalrez, SS316
Communication Protocol: RS-485 MODBUS, 4-20mA, Ethernet TCP
Power: 24VDC
Dimensions: 9.00″ (w) x 8.00″ (h) x 11.00″ (d) [22.86 x 20.32 x 27.94 cm]
Weight: 9.3 lbs (6.5 kg)

Conditions

Conditions

Operating: 50°F to 104°F (10°C to 40°C) / 20% to 95% non-condensing
Storage: 14°F to 122°F (-10°C to 50°C) / Up to 98% non-condensing

Accessories

Accessories

 

Included: Operating Manual on CD; Connection Tubing; Flow Cell Cleaning Solution and Brush; Carrying Case; 2 each Industrial RJ45 kit; 24VDC Power Supply and Power Cord; Communication Cable
Optional: Printed Operating Manual; Additional Tubing